Precision Glass Capillary Viscomters

ASTM D446 and ISO 3105
Standard Specifications and Operating Instructions for Glass Capillary Kinematic Viscometer

Cannon-Fenske Routine Viscometer For Transparent Liquds  ASTM D446 and ISO 3105

This is the most widely used type of glass capillary viscometer for measurements of fluids with Newtonian flow behavior as described in ASTM D445 and D446. Approximately 7ml of sample is required to fill the tube. To fill, the tube is inverted, and sample is suctioned into the double bulb section to the final timing mark. After tube is returned to upright position and sample has descended into the reservoir, the unit is equilibrated in bath at 40° C for 15-20 minutes. Viscosity of the sample is determined by pressuring the sample upward, timing the flow between timing marks on descent, and multiplying this time (in seconds) by the Constant of that specific viscometer. Successive timings may be done to confirm accuracy of the timing. Each viscometer has a permanent serial number and is supplied with a calibration certificate indicating the Constant of that tube at 40° C and 100° C.

Bath depth required: 8”