OilLab 880 Saybolt Viscometer

OilLab 880 Saybolt Viscometer

Download Brochure

ASTM D88
ASTM D7496
ASTM E102
IP 55
FTM 791-0304
JIS K 2207

CE marked

Subject

ASTM D88 Saybolt Viscosity: covers the measurement of viscosities of petroleum products at temperature between 21° and 99°C (70° ÷ 210°F)
ASTM D7496: This test method utilizes the Saybolt Furol viscometer to measure the consistency of emulsified asphalt. It is applicable to all the emulsified asphalts specified in Specifications D977 and D2397.
ASTM E 102 Saybolt Viscosity: covers the measurement of viscosities of petroleum products at temperature between 121° and 232°C (250° ÷ 450°F)

Principle of detection

The sample is heated and mantained stable at test temperature. Once the analisys start the valve is opened and a real time CCD camera will observe the flow of the sample until reach the level mark and close the valve. The time and relevant viscosity index (SUS and/or SFS) are automatically calculated.

Measuring Temperature Devices

  • Bath temperature: PT100 sensor stainless and steel

Detection system

  • Integrate CMOS sensor high definition
  • Integrate LED backlight

Heater

  • Electrical heating

Safety Devices

  • Safety thermostat over-temperature cut-out

Software Features

  • User friendly interface
  • All analytical parameters recorded
  • Customizable analysis parameters and methods
  • Customizable results report
  • Printable graphs and results
  • Orifices calibration procedure with reference sample
  • Centering procedure of the level mark of the receiver flask
  • Automatic open/close valves
  • Automatic viscosity calculation
  • Audible chimie alarm for over-temperature
  • Selectable temperature displaying: °C / °F

The software includes:

Analysis Menu

  • Automatic execution of the analysis in accordance to the selected procedure (Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference)
  • Display in real time of all the analysis parameters and status
  • Fields for introduction of operator and product name
  • Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions

Diagnostic Menu

  • Direct access to all analog, digital, inputs and outputs
  • Selectable value displaying: °C / Volt

Calibration Menu

  • Up to 100 calibration points (standard with 5 and dynamic with up to 100 points)
  • Last calibration date referred to each single probe displayed and relevant data printable
  • Display of calibration diagram
  • Comparative with reference thermometer
  • Selection calibration due date

Data Utilities

  • Archive viewer for files recall
  • All analysis stored in Excel® compatible format

Analysis storage

  • Storage capacity for more than 65.000 analysis
  • Export of test results files in the most common formats Excel and PDF
  • Reading interval of PT100 sample from 0 to 450 °C with resolution 0.1 °C

Integrated Touch Screen Panel PC

  • TFT/LCD 8.4″
  • Resolution 1024 × 768, 16.2 M colours
  • 1 × USB ports for connection to an external printer

Electrical Supply

  • 220V ± 15% / 50 to 60 Hz
  • 115V ± 15% / 60 Hz

Ambient Temperature

  • Max 35 °C
  • H.R. 80%

Weight

  • 45 kg

Dimensions

  • width: 43 cm
  • depth: 38 cm
  • height: 60 cm

Accessories

  • LAB-100-161: filter funnel with stainless steel wire mesh
  • LAB-100-162: Saybolt flask 60 ml
  • LAB-100-163: thermometer support
  • LAB-100-164: withdrawal tube
  • LAB-100-167: movement ring ASTM E 102
  • LAB-100-168: suction pipette

Spare Parts

  • LAB-100-161: filter funnel with stainless steel wire mesh
  • LAB-100-162: Saybolt flask
  • LAB-100-164: withdrawal tube
  • LAB-100-165: universal orifice
  • LAB-100-166: furol orifice
  • LAB-100-168: suction pipette
  • LAB-100-371: silicone oil 25 litres
  • LAB-140-001/A: PT100 probe
  • LAB-11-0012: heater