NewLab 410 Freezing Point |
Download Brochure |
ASTM D852 – ASTM D1177 – ASTM D1493 – ASTM D1655 – ASTM D2386
ASTM D5901 – ASTM D5972 (correlated) – ASTM D7153 (correlated)
ASTM D7154 (correlated)
IP 16 – IP 435 – IP 528 – IP 529
ISO 3013
CE marked
Subject
Freezing Point of aviation fuels, aviation gasoline, aviation turbine fuels, engine coolants, antifreeze products, brake fluids, …
Solidification Point of Benzene.
Solidification Point of Industrial Organic Chemicals.
Measuring Freezing Point Principle
According to the methods, the sample is cooled down and stirred. The solid hydrocarbon crystals formation are detected by means of a light beam throught fiber optic reflected thanks to a mirror. As soon as crystals are detected, the sample is warmed up until their complete disappearance.
Measuring Freezing Point Devices
- Two PT100 detection probes placed on the surface of the product
- Mechanical moving arm bringing the test jar in horizontal position
Measuring Temperature Probe
- Platinum resistance PT100 class A
Stirrer
- A micro-motor drives all the mechanical system
- 3 coils stirrer made of brass
Measuring Parameters
- Temperatures: in °C / °F
- Measuring range: -110°C … +100°C
- Range of analysis: -90°C … +55°C (410/2-SA)
- Resolution: 0.01 °C
- Accuracy: ± 0.1 °C
- Repeatability / Reproducibility: as per standards methods or better
Software Features
New LabLink software able to manage up to 6 analytical heads simultaneously (stand alone)
- User friendly interface
- All analytical parameters recorded
- Customizable analysis parameters and methods
- Results report
- Printable graphs and results – any Windows® compatible printer can be used
The software includes:
Analysis Menu
- Standard method as per ASTM / IP / ISO / EN / DIN… norms of reference:
- Optional methods:
- special detection of contaminants
- Audible alarm and displayed messages at the end of the analysis and in case of errors and/or malfunctions
Diagnostic Menu
- Direct access to all analog, digital, inputs and outputs
- Selectable value displaying: °C / Volt
Calibration Menu
- Automatic calibration of each temperature probe
- Last calibration date referred to each single probe displayed and relative data printable
- Display of calibration diagram
- Insertion of offset values
- Standard and advanced calibration modes up to 100 calibration points
Data Utilities
- Fields for introduction of operator and product name
- Archive viewer for files recall
- All analysis stored in Excel® compatible format and JPG image
- Storage capacity for more than 60’000 analysis
- LIMS compatible
Integrated panel PC
- TFT/LCD 12″
- Resolution 1024 × 768, 16.2 M colours
- 2 USB ports for connection to an external printer and/or external PC
- Storage capacity for more than 60’000 analysis
Test Jar
- Same dimensions and volume as described by the standard test methods
- Product level mark at 25 ml
- Small edge on the top in order to fix the glass cell to the analytical head
Cooling System
Insulated cooling jackets
Integrated gas CFC free motor compressors:
- Single stage (for temperatures up to -40°C / 1)
- Double stage (for temperatures up to -80°C / 2)
Equipped with an automatic energy power save system. After 15 minutes from the end of the analysis the cooling system goes in stand-by mode.
Safety Devices
- Pressure controller for 1st stage motor compressor
- Pressure controller for 2nd stage motor compressor
- Thermostat for 2nd stage activation
- Thermo-switch for each cooling / heating jacket
- Motor compressors equipped with internal overload devices
Electrical Supply
- 220V ± 15% / 50 to 60 Hz
- 115V ± 15% / 60 Hz
Cord Cable:
- 3 conductors flexible cable 2 m (7 feet) length with PVC sheath oil and heat resistant as per CENELEC directives
Ambient Temperature
- Max 32 °C
- H.R. 80%
Dimensions (cm) | 1 test position | 2 test position | 3 test position | 4 test position | 6 test position |
width | 66 | 66 | 100 | 134 | 130 |
depth | 60 | 60 | 60 | 60 | 75 |
height | 80 | 80 | 80 | 80 | 170 |
Weight (kg) | |||||
single stage compressor | 70 | 90 | – | – | – |
double stage compressor | 80 | 100 | 130 | 160 | 280 |
NewLab 410 ST
- Measuring range: -110°C …+100°C
- Range of analysis: -110°C…+55°C
- Resolution: 0.01 °C
- Width: 34 cm
- Depth: 60 cm
- Height: 80 cm
- Weight: 34 kg
Spare Parts
- LAB-xxx/005-03: heather + auto adhesive + insulation
- LAB-xxx/005-04: thermo switch
- LAB-xxx/005-06: PT100 bath
- LAB-xxx/007-02: static relay
- LAB-xxx/007-04: PCB fuse 1 AT (box of 10 pcs)
- LAB-xxx/006-01: cooling fluid valve + fitting
- LAB-400/007-01: main electronic board Freezing Point
- LAB-400/008-04: PT100 product w/connector
- LAB-400/008-05: stirrer
- LAB-400/008-08: mirror for Freezing Point
- LAB-400/008-06: motor for stirrer
- LAB-400/008-07: fibber optic for Freezing Point
- LAB-400/008-09: electronic board for detection
- LAB-410/008-12: removable glass cell for Freezing Point
- LAB-410/008-041: o-ring for Freezing Point test jar
Calibration Tools
- OilLab 80: calibration decade box – PT100 simulator
- OilLab 81: set of connectors and cables for cold range